SumMet Method for Chromium

SUMMET METHOD FOR CHROMIUM

Surface Abrasive / Size Load - lbs [N] / Specimen Base Speed [rpm] Relative Rotation Time [min:sec]
CarbiMet 320 [P400] grit SiC water cooled 6 [27] 300 Rotation Image Until Plane
UltraPad 9 µm MetaDi Supreme Diamond 6 [27] 150 Rotation Image 10:00
TriDent 3 µm MetaDi Supreme Diamond 6 [27] 150 Rotation Image 8:00